Models and specifications
Values are taken from the ClearPeak Raman System datasheet. For a specification not listed here, ask an applications engineer.
| Detector | Excitation (nm) | Laser (mW) |
|---|---|---|
| Sony IMX428 CMOS (-5 C to -10 C) | 830 | 450 |
- Primary laser wavelength 830 nm; 785 nm and 532 nm also available. Laser power at the sample is software-controlled, up to 450 mW. Raman shift range 100–1800 cm⁻¹.
A dash means the value is not specified for that model. An applications engineer can help with any specification not listed.
Common to every model
- Power
- 100-240 V AC single-phase, 50-60 Hz, 37 W (max)
- Raman shift range (cm⁻¹)
- 100-1800
- Étendue
- 6.5e-3 cm^2 sr
- Narrow FOV interrogation diameter (mm)
- 1.4
- Wide FOV interrogation diameter (mm)
- 14
- Narrow FOV illumination diameter (mm)
- 0.25
- Wide FOV illumination diameter (mm)
- 2
- Narrow FOV working distance (mm)
- 12.5
- Wide FOV working distance (mm)
- 50
- Typical integration time
- 1 s to 60 s (depends on chemical target)
- SNR (liquid)
- 1500 (acetonitrile in quartz cuvette)
- SNR (solid)
- 1000 (polystyrene)
- Detector cooling
- -5 C to -10 C
- Weight (lb)
- 12
- Dimensions, main body
- 14 x 11 x 4.5 cu. in.
- Dimensions, fiber attachment
- 5.9 x 6.4 x 4.5 cu. in.
- Laser safety class
- 3b