DragonFly NIR Spectrometer
Programmable NIR based on Texas Instruments' DLP technology
900–2280 nm10 nm resolutionSNR to 70000.08 nm/°CTEC option4variants
Used in Pharmaceutical, Industrial & materials
Models and specifications
Values are taken from the DragonFly NIR Spectrometer datasheet. For a specification not listed here, ask an applications engineer.
| Model | Detector | Range (nm) | Resolution (nm) | SNR | Stray light (%) | Dark noise | Cooling |
|---|---|---|---|---|---|---|---|
| DF1514 1 | InGaAs PIN photodiode, Ø1 mm (Hamamatsu G12180 series) | 900–1700 | 10 | 7000 | 0.2 | 15 | None |
| DF1914 2 | InGaAs PIN photodiode, Ø1 mm (Hamamatsu G12180 series) | 1250–2050 | 11 | 7000 | 0.2 | 15 | TEC, sensor held near −20 °C |
| DF1924 3 | InGaAs PIN photodiode, Ø1 mm (Hamamatsu G12180 series) | 1340–2280 | 13 | 7000 | 0.2 | 15 | TEC, sensor held near −20 °C |
| DF1934 4 | — | — | — | — | — | — | TEC, sensor held near −20 °C |
- DF1514: Dimensions 71.5 × 57 × 25 mm; weight 77 g. 24-bit A/D. Slit width 25 µm. Minimum exposure time 0.635 ms.
- DF1914: Dimensions 76.7 × 59.9 × 40 mm; weight 120 g. 24-bit A/D.
- DF1924: Dimensions 76.7 × 59.9 × 40 mm; weight 120 g. 24-bit A/D.
- DF1934: Dimensions 76.7 × 60 × 40 mm.
A dash means the value is not specified for that model. An applications engineer can help with any specification not listed.
Common to every model
- A/D converter
- 24-bit
- Interface
- Micro USB 2.0 and UART (GPIO)
- Connector
- SMA905 (Ø 3.18 ± 0.005 mm ferrule, fiber input)
- Thermal stability (nm/°C)
- < 0.08 nm/°C
- Operating temperature (°C)
- 0 – 40 °C
- Storage temperature (°C)
- −30 to 70 °C
- Humidity
- 0–85% RH
- Power
- 5 V DC, 300 mA typical; 500 mA maximum with TEC cooling
- Slits
- 25 µm (fixed)
- DMD resolution
- 854 × 480 micro-mirrors (Texas Instruments DLP design)
- Wavelength accuracy
- ±1 nm
- Minimum exposure time
- 0.635 ms
- Hadamard scan mode
- Hadamard scan mode improves SNR by up to 2–7× over standard column scanning