Contact an engineer

System

Dual Spectrometer Systems (DS Series)

Integrate two spectrometers with a common trigger for extended range

180–1700 nm11variants

Used in Semiconductor, Environmental, Industrial & materials

Dual Spectrometer Systems (DS Series)

Models and specifications

Values are taken from the Dual Spectrometer Systems (DS Series) datasheet. For a specification not listed here, ask an applications engineer.

Models and specifications: System
ModelDetectorRange (nm)Connector
SE2030 1 2048-pixel CMOS 180–1025 SMA905
SE2040 2 2048-pixel CCD 275–1100 SMA905
SE2050 3 2048-pixel CCD 350–1020 SMA905
SE2060 4 2048-pixel CCD 350–1025 SMA905
SE2080 5 4096-pixel CMOS 350–1010 SMA905
EE2053 6 TEC backthinned CCD 180–1025 SMA905
EE2063 7 TEC backthinned CCD 350–1025 SMA905
SW2520 8 128-pixel InGaAs 900–1700 SMA905
SW2530 9 256-pixel InGaAs 900–1700 SMA905
SW2540 10 512-pixel InGaAs 900–1700 SMA905
SW2860 11 256-pixel InGaAs (TEC) 900–1700 SMA905
  1. SE2030: SNR 330; dynamic range 3450; 16-bit A/D; stray light 0.2%; thermal stability <0.04 nm/°C.
  2. SE2040: SNR 200; dynamic range 3000; 16-bit A/D; stray light 0.2%; thermal stability <0.04 nm/°C.
  3. SE2050: SNR 500; dynamic range 2200; 16-bit A/D; stray light 0.45%; thermal stability <0.04 nm/°C.
  4. SE2060: SNR 500; dynamic range 4700; 16-bit A/D; stray light 0.2%; thermal stability <0.04 nm/°C.
  5. SE2080: SNR 350; dynamic range 2200; 16-bit A/D; stray light 0.2%; thermal stability <0.04 nm/°C.
  6. EE2053: SNR 500; dynamic range 4700; 16-bit A/D; stray light 0.45%; thermal stability <0.04 nm/°C.
  7. EE2063: SNR 500; dynamic range 4096; 16-bit A/D; stray light 0.45%; thermal stability <0.04 nm/°C.
  8. SW2520: SNR 4000 / 2000 (low / high gain); dynamic range 6000 / 4100; 16-bit A/D; stray light 0.2%; thermal stability <0.069 nm/°C.
  9. SW2530: SNR 4000 / 2000 (low / high gain); dynamic range 6000 / 4100; 16-bit A/D; stray light 0.2%; thermal stability <0.069 nm/°C.
  10. SW2540: SNR 4000 / 2000 (low / high gain); dynamic range 6000 / 4100; 16-bit A/D; stray light 0.2%; thermal stability <0.069 nm/°C.
  11. SW2860: SNR 4100 / 2700 (low / high gain); dynamic range 5242 / 4854; 16-bit A/D; stray light 0.2%; thermal stability <0.069 nm/°C.

A dash means the value is not specified for that model. An applications engineer can help with any specification not listed.

Common to every model

Interface
USB 2.0 high-speed (480 Mbps), PC-controlled with SDK support
Dimensions (mm)
180 (L) x 175 (W) x 60.7 (H) mm (standard)
Power
USB: ~300 mA @ +5 VDC; TEC option: 500 mA @ +5 VDC per unit
Operating temperature (°C)
0-50 C operation; -30-70 C storage; 0-90% RH non-condensing
Optical resolution
0.2-10.5 nm (depending on sensor/slit)
Integration time
0.2 ms - 65 s (UV-VIS); 50 us - 15 s (NIR)
Slits, UV-Vis channel (µm)
10, 25, 50, 100, 200, 300
Slits, NIR channel (µm)
10-200
Fiber connector
SMA905 (IEC 874-2 standard, ferrule <9.812 mm)
Example pairing
DS2686 = EE2060 + SW2860
Overall range
180–1700 nm overall, with seamless overlap between the VIS and NIR channels

Next step

Discuss the Dual Spectrometer Systems (DS Series) with an engineer

Send the sample type, the wavelength range and the accuracy required. An Evolve Sensing engineer will confirm the configuration, or recommend a different instrument where one suits the measurement better.