180–1100 nm0.2 nm resolutionSNR to 5000.015 nm/°C2variants

Used in Biomedical, Pharmaceutical, Environmental

EagleEye UV-Vis-NIR Spectrometer

Models and specifications

Values are taken from the EagleEye UV-Vis-NIR Spectrometer datasheet. For a specification not listed here, ask an applications engineer.

Models and specifications: UV-Vis
ModelDetectorPixelsRange (nm)Resolution (nm)SNRDyn. rangeStray light (%)Dark noise
EE2113 1 NIR Enhanced BackThinned TEC linear sensor (Hamamatsu S16011) 2048 500–1100 0.2–10.5 500 5000 0.45 13
EE2063 2 UV Enhanced low-noise Back-Thinned TEC linear sensor (Hamamatsu S11850) 2048 180–1100 0.2–10.5 500 4000 0.45 17
  1. EE2113: Thermal stability <0.015 nm/°C.
  2. EE2063: Thermal stability <0.04 nm/°C.

A dash means the value is not specified for that model. An applications engineer can help with any specification not listed.

Common to every model

Integration time
5 ms - 65 s
A/D converter
16
Interface
USB 2.0 @ 480 Mbps
Thermal stability (nm/°C)
model dependent: <0.015 (EE2113), <0.04 (EE2063)
Operating temperature (°C)
0 to 50
Storage temperature (°C)
-30 to 70
Humidity
0-90% non-condensing
Dimensions (mm)
130(L) x 96(W) x 58.3(H)
Weight (kg)
0.88
Power
USB: 330 mA @ +5 VDC; TEC: 500 mA @ +5 VDC; voltage range 4.75-5.25 V
Gratings
15 grating options, spectral range from UV to NIR
Slits
10, 25, 50, 100, 200, 300 µm
Optical design
Czerny-Turner, 2nd & 3rd order rejection
Wavelength accuracy
±0.3 nm typical (±1.0 nm extremes)
Wavelength repeatability
±0.05 nm
Cooling
20-25°C below ambient
Connectors
8-pin I/O (6 GPIO, TX/RX, Trigger); SMA905 fiber connector (Ø3.18 ± 0.005 mm ferrule)