Market-leading NIR performance

900–1700 nmSNR to 30002variants

Used in Pharmaceutical, Industrial & materials

RedBullet NIR Spectrometer

Models and specifications

Values are taken from the RedBullet NIR Spectrometer datasheet. For a specification not listed here, ask an applications engineer.

Models and specifications: NIR
ModelDetectorPixelsRange (nm)Resolution (nm)SNRDyn. rangeStray light (%)Dark noise
RB4524 1 InGaAs (uncooled) 128 900–1700 <15 2500 6500 0.15 10
RB4564 2 InGaAs (uncooled) 256 900–1700 <12 3000 7300 0.15 9
  1. RB4524: Optical resolution <15 nm at 50 µm slit. Dimensions 50 × 36.4 × 28.4 mm; weight 60 g.
  2. RB4564: Optical resolution <12 nm at 50 µm slit. Dimensions 50 × 36.4 × 28.4 mm; weight 60 g.

A dash means the value is not specified for that model. An applications engineer can help with any specification not listed.

Common to every model

Integration time
6 µs – 24 s
A/D converter
16-bit, 15 MHz
Interface
USB 2.0 @ 480 Mbps (Micro USB or 4-pin USB); programmable digital I/O (six 3.3V GPIO, UART TX/RX, external trigger, lamp control)
Connector
SMA905 (IEC 874-2:1993 standard, fiber); Micro USB / 4-pin USB
Operating temperature (°C)
0 – 50 °C
Storage temperature (°C)
−30 to 70 °C
Humidity
0–90% non-condensing
Weight (kg)
0.06
Power
500 mA at +5 V DC (4.75–5.25 V)
Gratings
Various grating options
Slits
25 / 50 / 100 µm (customizable)
Optical design
Concave mirror Czerny–Turner, 2nd & 3rd harmonics removed (f/4.5, NA 0.11)
Baseline signal
1,000 counts (adjustable via software)