Contact an engineer

NIR

SideWinder NIR Spectrometer

Market-leading NIR performance

900–2500 nmSNR to 32000.1 nm/°CTEC option8variants

Used in Pharmaceutical, Environmental, Industrial & materials

SideWinder NIR Spectrometer

Models and specifications

Values are taken from the SideWinder NIR Spectrometer datasheet. For a specification not listed here, ask an applications engineer.

Models and specifications: NIR
ModelDetectorPixelsRange (nm)SNRDyn. rangeDark noiseCooling
SW2520 1 InGaAs (uncooled) 128 900–1700 2000 6500 10 Uncooled
SW2560 2 InGaAs (TEC, 1-stage) 256 910–2200 2400 6000 11 —
SW2570 3 InGaAs (TEC, 2-stage) 512 900–2500 2500 8000 11 —
SW2860 4 InGaAs (TEC, 1-stage) 256 910–2200 3200 7200 8 TEC, 1-stage
SW2870 5 InGaAs (TEC, 2-stage) 512 900–2500 2800 6000 9 TEC, 2-stage
SW2930 6 InGaAs (TEC, 1-stage) 256 910–2200 2500 7000 11 TEC, 1-stage
SW2960 7 InGaAs (TEC, 1-stage) 256 900–2500 — 6000 8.5 TEC, 1-stage
SW2970 8 InGaAs (TEC, 2-stage) 512 900–2500 — — 11 TEC, 2-stage
  1. SW2520: Integration time 100 µs – 24 s. Dimensions 110 × 86 × 32.4 mm.
  2. SW2560: Integration time 100 µs – 300 ms.
  3. SW2570: Integration time 100 µs – 200 ms.
  4. SW2860: Integration time 100 µs – 300 ms. Dimensions 130 × 96 × 57.65 mm.
  5. SW2870: Integration time 100 µs – 200 ms. Dimensions 130 × 96 × 57.65 mm.
  6. SW2930: Integration time 100 µs – 300 ms. Dimensions 130 × 96 × 57.65 mm.
  7. SW2960: Integration time 100 µs – 200 ms.
  8. SW2970: Integration time 100 µs – 200 ms.

A dash means the value is not specified for that model. An applications engineer can help with any specification not listed.

Common to every model

A/D converter
16-bit, 15 MHz
Interface
USB 2.0 @ 480 Mbps; 6× programmable 3.3V GPIO
Connector
SMA905 (fiber)
Thermal stability (nm/°C)
< 0.1 – 0.2 nm/°C (model dependent)
Operating temperature (°C)
0 – 50 °C
Storage temperature (°C)
−30 to 70 °C
Humidity
0–90% non-condensing
Weight (kg)
0.9
Power
SW2520/2560/2570: +5 V DC, 300–350 mA (USB). SW2860/2870: +5 V DC, 350 mA (USB) + 1300 mA (TEC). SW2930/2960/2970: +5 V DC, 480 mA (USB) + 2200 mA (TEC).
Slits
25–200 µm (model dependent)
Optical design
Czerny–Turner, 2nd & 3rd order rejection; f/#5, NA 0.1
Wavelength accuracy
±1 nm typical; up to ±1.5 nm in extreme conditions
Repeatability
±0.2 nm (100 continuous measurements, Ar light source)
Resolution range
2–22 nm (depending on slit, sensor and grating)
Stray light
<0.1–0.45%, depending on model and wavelength

Next step

Discuss the SideWinder NIR Spectrometer with an engineer

Send the sample type, the wavelength range and the accuracy required. An Evolve Sensing engineer will confirm the configuration, or recommend a different instrument where one suits the measurement better.