Contact an engineer

UV-Vis

SilverBullet Compact OEM UV-Vis Spectrometer

Market-leading UV-Vis performance

3 nm resolutionSNR to 3500.04 nm/°C2variants

Used in Semiconductor, Pharmaceutical

SilverBullet Compact OEM UV-Vis Spectrometer

Models and specifications

Values are taken from the SilverBullet Compact OEM UV-Vis Spectrometer datasheet. For a specification not listed here, ask an applications engineer.

Models and specifications: UV-Vis
ModelDetectorPixelsResolution (nm)SNRDyn. rangeStray light (%)Dark noise
SB2134/SB3134/SB4134 1 CMOS 1024 3–15 350 5200 0.3 12.5
SB2130/SB3130/SB4130 2 CMOS 1024 3–15 350 5200 0.3 12.5
  1. SB2134/SB3134/SB4134: Wavelength range depends on the grating option: FUVA 180–850 nm, DUV2B 200–850 nm or VNIR7B 300–1100 nm. Dimensions 40 × 36.3 × 25.1 mm (L × W × H).
  2. SB2130/SB3130/SB4130: Wavelength range depends on the grating option: FUVA 180–850 nm, DUV2B 200–850 nm or VNIR7B 300–1100 nm. Dimensions 49 × 48 × 28.5 mm (L × W × H).

A dash means the value is not specified for that model. An applications engineer can help with any specification not listed.

Common to every model

Integration time
6 µs - 65 s (10 MHz); 21 µs - 65 s (5 MHz)
A/D converter
16
Interface
USB 2.0 (480 Mbit/s), programmable digital I/O
Thermal stability (nm/°C)
<0.04
Operating temperature (°C)
0 to 50
Storage temperature (°C)
-30 to 70
Humidity
0-90% non-condensing
Dimensions (mm)
SB3134/4134: 40(L) x 36.3(W) x 25.1(H); SB3130/4130: 49(L) x 48(W) x 28.5(H)
Weight (kg)
0.036
Power
500 mA @ +5 VDC (4.75-5.25 V range)
Connectors
SMA905 fiber optic; 8-pin rear I/O port (6 GPIO, 1 trigger, 1 lamp control)

Next step

Discuss the SilverBullet Compact OEM UV-Vis Spectrometer with an engineer

Send the sample type, the wavelength range and the accuracy required. An Evolve Sensing engineer will confirm the configuration, or recommend a different instrument where one suits the measurement better.