Contact an engineer

UV-Vis

SmartEngine UV-VNIR Spectrometer

Workhorse UV-VNIR platform with CCD and CMOS sensor options

180–1100 nm0.2 nm resolutionSNR to 5000.04 nm/°C9variants

Used in Semiconductor, Pharmaceutical, Environmental, Industrial & materials

SmartEngine UV-VNIR Spectrometer

Models and specifications

Values are taken from the SmartEngine UV-VNIR Spectrometer datasheet. For a specification not listed here, ask an applications engineer.

Models and specifications: UV-Vis
ModelDetectorPixelsRange (nm)Resolution (nm)SNRDyn. rangeStray light (%)Dark noise
SE1010/SE2010 Front illuminated CCD 3000 180–1100 0.2–10.5 200 2200 0.2 30
SE1020/SE2020 UV enhanced front illuminated CCD 2048 180–1100 0.2–10.5 250 1700 0.45 38
SE1030/SE2030 1 Fast exposure CMOS 2048 180–1100 0.2–10.5 350 — 0.2 —
SE1040/SE2040 Front illuminated CCD 3648 180–1100 0.2–10.5 200 2200 0.45 30
SE1050/SE2050 NIR enhanced back illuminated CCD 4096 180–1100 0.2–10.5 500 4600 0.2 14
SE1070/SE2070 High pixel-resolution CCD 2048 180–1100 0.2–10.5 350 2200 0.45 30
SE1080/SE2080 Fast exposure CMOS 1024 180–1100 0.2–10.5 500 3200 0.2 21
SE1090/SE2090/SE5090/SE6090/SE5190/SE6190 Fast exposure back illuminated CMOS 1024 180–1100 0.2–10.5 350 6000 — 11
SE1120/SE2120 Fast exposure CMOS 1024 180–1100 0.2–10.5 350 4300 — 15
  1. SE1030/SE2030: Dynamic range 4300 at 2.5 MHz, 3000 at 10 MHz; dark noise 15 at 2.5 MHz, 21 at 10 MHz.

A dash means the value is not specified for that model. An applications engineer can help with any specification not listed.

Common to every model

Integration time
0.1 ms - 65 s
A/D converter
16
Interface
USB 2.0 (480 Mbps) + programmable I/O
Thermal stability (nm/°C)
<0.04
Operating temperature (°C)
0 to 50
Storage temperature (°C)
-30 to 70
Humidity
0-90% non-condensing
Dimensions (mm)
Type I: 110(L) x 86(W) x 32.4(H); Type II: 110(L) x 86(W) x 35.4(H); Double-layer: 110(L) x 86(W) x 53.9(H)
Weight (kg)
0.4
Power
+5 V DC via USB, 300 mA typical
Gratings
15 choices, from UV to NIR
Boot time
<4 seconds
Optical system
Czerny-Turner, f/# 5, NA 0.1, 60 mm focal length
Wavelength accuracy
±0.3 nm typical (±1 nm extreme conditions)
Wavelength reproducibility
±0.05 nm
Connectors
8-pin, with 6 programmable GPIO lines